临床研究

镍钛记忆合金腕骨四角融合器应用于腕塌陷临床疗效:9~10年随访

展开
  • 1.解放军联勤保障部队第九二〇医院骨科,  昆明   650032;    2.昆明医科大学研究生院,  昆明   650500
齐保闯(1980-),男,山东聊城人,博士研究生,副主任医师,研究方向:关节外科,E-mail:qibaochuang@126.com
徐永清,教授,主任医师,博士研究生导师,E-mail:xuyongqingkm@163.net

收稿日期: 2023-11-07

  网络出版日期: 2024-06-28

基金资助

云南省骨科与运动康复临床医学研究中心(202102 AA310068);云南省创伤骨科临床医学中心(ZX20191001);全军临床重点专科建设项目;联勤医学重点专科项目

Clinical efficacy of four-corner arthrodesis concentrator of nickel-titanium memory alloy for carpal collapse: 9-10 years of follow-up 

Expand
  • 1.Department of  Orthopedics, 920th Hospital of Joint Logistics Support Force of Chinese People's Liberation Army, Kunming 650032, China; 2. Graduate School of Kunming Medical University, Kunming 650500, China

Received date: 2023-11-07

  Online published: 2024-06-28

摘要

目的    探讨镍钛记忆合金腕骨四角融合器(NiTi shape memory ally four-corner arthrodesis concentrator,NT-FCAC)在治疗腕塌陷的应用效果,报告9~10年随访疗效。  方法    选取32例因舟骨骨不连腕塌陷及舟月进行性腕塌陷应用NT-FCAC行四角融合并舟骨切除术治疗的患者,采用测角计测量腕关节活动度;Jammar测力计测量握力;术前及术后进行快速臂肩手功能障碍评分(Quick DASH)观察肢体功能恢复情况;疼痛视觉模拟评分(VAS)评估患者腕关节疼痛程度,腕关节前后位及侧位片观察腕骨融合率。  结果    术后3个月32例腕骨均获骨性愈合。随访9~10年,所有患者未出现严重手术并发症,腕关节活动范围达到健侧的60%以上,患手握力恢复至健侧的80.78%。VAS评分由术前(5.97±0.86)分减至末次随访(0.16±0.37)分;Quick DASH评分由术前(70.00±4.76)分降至末次随访(6.16±1.25)分。VAS及Quick DASH评分术前较术后各次随访均具有统计学差异(P<0.05)。  结论    应用NT-FCAC进行腕骨四角融合术,腕骨融合率高,能保存大部分腕关节功能,远期疗效好,适用于需四角融合的腕塌陷。

本文引用格式

齐保闯, 蔡兴博, 宋慕国, 孟晨, 罗欢, 李川, 徐永清 . 镍钛记忆合金腕骨四角融合器应用于腕塌陷临床疗效:9~10年随访[J]. 中国临床解剖学杂志, 2024 , 42(3) : 304 -309 . DOI: 10.13418/j.issn.1001-165x.2024.3.10

Abstract

 Objective   To explore the application effect of NiTi shape memory ally four-corner arthrodesis concentrator (NT-FCAC) in the treatment of wrist collapse, and to report the efficacy of 9-10 years follow-up.    Methods    Thirty-two patients with four-corner arthrodesis and navicular osteotomy with NT-FCAC were selected for treatment of wrist collapse due to navicular osteogenesis imperfecta and navicular-moon-progressive wrist collapse. Goniometer was used to measure the wrist joint mobility. Jammar dynamometer was used to measure the grip strength. Quick DASH (Quick Arm, Shoulder, Hand Dysfunction Score) was used to observe the recovery of the limb function before and after the operation, and Visual Analogue Score (VAS) was used to assess the degree of the patients' wrist pain. Anteroposterior and lateral radiographs of the wrist joint were taken to observe the rate of carpal arthrodesis.    Results   Bone healing was obtained in all 32 carpal bones at 3 months postoperatively. At 9-10 years of follow-up, all patients had no serious surgical complications, the range of motion of the wrist joint reached more than 60% of the healthy side, and the grip strength of the affected hand was restored to 80.78% of that of the healthy side. The VAS score was reduced from the preoperative score (5.97±0.86) to the final follow-up (0.16±0.37), the Quick DASH score was reduced from the preoperative score (70.00±4.76) to the final follow-up (6.16±0.37). There were statistical differences in the preoperative  VAS and Quick DASH scores and  the VAS and Quick DASH scores at each follow-up.    Conclusions    The application of NT-FCAC for four-corner arthrodesis of the carpal bones has a high rate of carpal fusion, preserves most of the functions of the carpal joints, and has a good long-term efficacy, which is suitable for carpal collapse that requires four-corner arthrodesis.

参考文献

[1]  Li C, Peng Z, Zhou Y, et al. The progress in the classification and treatment of scaphoid nonunion[J]. Surgeon, 2022, 20(5): e231-e235. DOI: 10.1016/j.surge.2022.05.006.
[2]  Oh WT, Park HJ, Koh IH, et al. Outcomes of arthroscopic scaphoid excision and lunocapitate fusion for advanced traumatic arthritis of the wrist[J]. Clin Orthop Surg, 2023, 15(2): 308-317. DOI: 10.4055/cios22066.
[3]  Watson HK, Goodman ML, Johnson TR. Limited wrist arthrodesis. Part II: intercarpal and radiocarpal combinations[J]. J Hand Surg Am, 1981, 6: 223-233. DOI: 10.1016/s0363-5023(81)80074-3.
[4] d'Almeida MA, Sturbois-Nachef N, Amouyel T, et al. Four-corner fusion: clinical and radiological outcome after fixation by headless compression screws or dorsal locking plate at minimum 5 years' follow-up[J]. Orthop Traumatol Surg Res, 2021, 107(5): 102886. DOI: 10.1016/j.otsr.2021.102886.
[5]  Reigstad O, Holm-Glad T, Dovland P, et al. Progressing arthrosis and a high conversion rate 11 (4-19) years after four corner fusion[J]. J Plast Surg Hand Surg, 2021, 55(6): 354-360. DOI: 10.1080/2000656X.2021.1898970.
[6]  Shindle M, Burton K, Weiland A, et al. Complications of circular plate fixation for four-corner arthrodesis[J]. J Hand Surg Eur Vol, 2007, 32(1): 50e53. DOI: 10.1016/j.jhsb.2006.08.016.
[7]  Vance MC, Hernandez JD, DiDonna ML, et al. Complications and outcome of four-corner arthrodesis: circular plate fixation versus traditional techniques[J]. J Hand Surg Am, 2005, 30(6): 1122e1127. DOI: 10.1016/j.jhsa.2005.08.007.
[8] Kendall CB, Brown TR, Millon SJ, et al. Results of four-corner arthrodesis using dorsal circular plate fixation[J]. J Hand Surg Am, 2005, 30(5): 903e907. DOI: 10.1016/j.jhsa.2005.04.007.
[9]  Xu YQ, Qi BC, Fan XY, et al. Four-corner arthrodesis concentrator of nickel-titanium memory alloy for carpal collapse: a report on 18 cases[J]. J Hand Surg Am, 2012, 37(11): 2246-2251. DOI: 10.1016/j.jhsa.2012.07.040.
[10]Vender MI, Watson HK, Wiener BD, et al. Degenerative change in symptomatic scaphoid nonunion[J]. J Hand Surg Am, 1987, 12(4): 514-519. DOI: 10.1016/s0363-5023(87)80198-3.
[11]Watson HK, Ballet FL. The SLAC wrist: scapholunate advanced collapse pattern of degenerative arthritis[J]. J Hand Surg Am, 1984, 9(3): 358-365. DOI: 10.1016/s0363-5023(84)80223-3.
[12]Petermann-Rocha F, Gray SR, Forrest E, et al. Associations of muscle mass and grip strength with severe NAFLD: a prospective study of 333,295 UK Biobank participants[J]. J Hepatol, 2022, 76(5): 1021-1029. DOI: 10.1016/j.jhep.2022.01.010.
[13] Alnahdi AH. Validity and reliability of the Arabic quick disabilities of the arm, Shoulder and Hand (QuickDASH-Arabic)[J]. Musculoskelet Sci Pract, 2021, 53: 102372. DOI: 10.1016/j.msksp.2021.102372.
[14]Duraku LS, Hundepool CA, Hoogendam L, et al. Two-corner fusion or four-corner fusion of the wrist for midcarpal osteoarthritis? A multicenter prospective comparative cohort study[J]. Plast Reconstr Surg, 2022, 149(6):1130e-1139e. DOI: 10.1097/PRS.00000000000 09116.
[15]Ahmadi AR, Duraku LS, van der Oest MJW, et al. The never-ending battle between proximal row carpectomy and four corner arthrodesis: a systematic review and meta-analysis for the final verdict[J]. J Plast Reconstr Aesthet Surg, 2022, 75(2): 711-721. DOI: 10.1016/j.bjps.2021.09.076.
[16] Van Amerongen EA, Schuurman AH. Four-corner arthrodesis using the Quad memory staple[J]. J Hand Surg Eur Vol, 2009, 34(2): 252-255. DOI: 10.1177/1753193408094444.
[17] Erne HC, Broer PN, Weiss F, et al. Four-corner fusion: comparing outcomes of conventional K-wire-, locking plate-, and retrograde headless compression screw fixations[J]. J Plast Reconstr Aesthet Surg, 2019, 72(6): 909-917. DOI: 10.1016/j.bjps.2018.12.033.
[18] d'Almeida MA, Sturbois-Nachef N, Amouyel T. Four-corner fusion: clinical and radiological outcome after fixation by headless compression screws or dorsal locking plate at minimum 5 years' follow-up[J]. Orthop Traumatol Surg Res, 2021, 107(5): 102886. DOI: 10.1016/j.otsr.2021.102886.
[19] Zenke Y, Oshige T, Menuki K, et al. Four-corner fusion method using a bioabsorbable plate for scapholunate advanced collapse and scaphoid nonunion advanced collapse wrists: a case series study[J]. BMC Musculoskelet Disord, 2020, 21(1): 683. DOI: 10.1186/s12891-020-03709-0.
[20] Reissner L, Hensler S, Kluge S, et al. Treatment outcomes of 4-corner arthrodesis for patients with advanced carpal collapse: an average of 4 years' follow-up comparing 2 different plate types[J]. J Hand Surg Am, 2018, 43(5): 487.e1-487.e6. DOI: 10.1016/j.jhsa.2017.10.036.
[21] Ole Reigstad, Trygve Holm-Glad, Preben Dovland, et al. Progressing arthrosis and a high conversion rate 11 (4-19) years after four corner fusion[J]. J Plast Surg Hand Surg, 2021, 55(6): 354-360. DOI: 10.1080/2000656X.2021.1898970.
[22] Xu YQ, Qi BC, Fan XY, et al. Four-corner arthrodesis concentrator of nickel-titanium memory alloy for carpal collapse: a report on 18 cases[J]. J Hand Surg Am, 2012, 37(11): 2246-2251. DOI: 10.1016/j.jhsa.2012.07.040.
[23] Le Corre A, Ardouin L, Loubersac T, et al. Retrospective study of two fixation methods for 4-corner fusion: shape-memory staple vs. dorsal circular plate[J]. Chir Main, 2015, 34(6): 300-306. DOI: 10.1016/j.main.2015.08.008.
[24] Mevel G, De Geyer A, Kim W, et al. Four-corner arthrodesis using the 4Fusion® quadripodal shape memory staple: a long-term review[J]. J Hand Surg Eur Vol, 2022, 47(4): 387-392. DOI: 10.1177/17531934 211063614.
[25] Andronic O, Labèr R, Kriechling P, et al. Surgical fixation techniques in four-corner fusion of the wrist: a systematic review of 1103 cases[J]. J Plast Surg Hand Surg, 2023, 57(1-6): 29-37. DOI: 10.1080/2000656X.2022.2070182.