跟骨内侧定位载距突螺钉导向器的研制与初步评价

刘金伟, 朱朝晖, 张增方, 刘得恒, 杨斌, 郑良孝, 陈东亮, 宋晓峰, 刘鹏, 赛佳明, 丁自海

中国临床解剖学杂志 ›› 2019, Vol. 37 ›› Issue (6) : 691-696.

中国临床解剖学杂志 ›› 2019, Vol. 37 ›› Issue (6) : 691-696. DOI: 10.13418/j.issn.1001-165x.2019.06.016
临床研究

跟骨内侧定位载距突螺钉导向器的研制与初步评价

  • 刘金伟1, 朱朝晖1, 张增方1, 刘得恒1, 杨斌1, 郑良孝1
    陈东亮1, 宋晓峰1, 刘鹏1, 赛佳明1, 丁自海2
作者信息 +

Development and preliminary evaluation of the calcaneal sustentaculum tali director using the method of positioning from the medial of the calcaneus 

  • LIU Jin-wei1, ZHU Zhao-hui1, ZHANG Zeng-fang1, LIU De-heng1, YANG Bin1, ZHENG Liang-xiao1, CHEN Dong-liang1, SONG Xiao-feng1, LIU Peng1, SAI Jia-ming1, DING Zi-hai2
Author information +
文章历史 +

摘要

目的 探讨从跟骨内侧定位、经载距突置入导向针后,由跟骨外侧壁置入载距突螺钉的可行性及精确度,为临床手术方案设计及导向器械的研发改进提供参考依据。  方法 选取10具20侧成人尸体标本,在跟骨外侧做“L”形延长切口,暴露跟骨外侧结构及距下关节,采用自主研发的“跟骨内侧定位载距突螺钉导向器”辅助载距突螺钉置入。首先,从内侧将“定位针”置入距下中关节,将导向器“内侧臂”的“定位孔”套牢定位针,调整“内侧臂”使“导向通道”位于载距突中心,安装“外侧臂”与“手柄”固定。调整“外侧臂”位置,经“导向通道”从内向外穿入跟骨一枚“导向针”,C臂透视提示“导向针”位置理想后,自跟骨外侧壁沿“导向针”用空心钻钻孔,测深后拧入合适的载距突螺钉。最后行CT扫描,利用CT图像评估置入螺钉的精确度。过程中采集相关数据,评估导向器的功能。  结果 肉眼观察标本,所有的导向针位于载距突内,其中65%位于载距突中心,25%偏下方,10%偏后方。在内侧,75%的导向针穿过胫后肌腱,15%位于胫后肌和趾长屈肌腱之间,10%穿经趾长屈肌腱上1/3。在外侧,导向针出针点与跟骰关节的距离为(38.03±5.60)mm,与后关节面的距离为(15.01±3.38)mm,螺钉平均长度为(44.80±3.59)mm。CT扫描图像可见,所有螺钉位于载距突内,有2例穿透上方骨皮质,无螺钉穿出载距突下方、前方及后方骨皮质。  结论 采用跟骨内侧定位法,在自主设计的导向器引导下,可提高载距突螺钉置入的准确性,降低螺钉进入关节、跟骨内侧皮质等并发症。

Abstract

Objective  To evaluate the feasibility and accuracy of the guide needle inserted through the calcaneal sustentaculum tali by locating from inside, while putting the screw from the outside.  Methods  The L-shaped extended incision was simulated to expose the lateral calcaneal structure and subtalar joints on 10 adult cadaveric specimens (20 sides). The self-designed calcaneal sustentaculum tali director was adopted. First, placed the "positioning needle" from the inside into the subtalar joint, made the “positioning hole” of the “inner arm” entangled the positioning needle, then placed and adjusted the "inner arm of the director" so that the latter "guiding channel" was located in the center of the sustentaculum tali, installed  “the lateral arm” to fix “the handle”. Then, a "guide needle" was inserted to the calcaneal from inside to outside via the"guide channel", after C-arm fluoroscopy clarifiing the ideal position of the "guide needle", drilled with a hollow hole from the outside along the "guide needle", then screwed in the appropriate hollow pressure screw. Finally, the specimens were scanned under spiral CT and the accuracy of the insertion screw was evaluated using CT images. Collecting relevant data during the process to evaluate the function of the director.  Results  The guide needles of all the specimens were located in the sustentaculum tali, 65% of them were located in the center, 25% below and 10% behind. On the inside, 75% of the guide needles passed through the posterior tibia tendon, 15% lied between the posterior tibia tendon and the flexor digitorum longus tendon, and 10% passes through the upper 1/3 of the flexor digitorum longus tendon. On the outside, the distance between the needle outlet point of the guide needle and the calcaneocuboid joint was (38.03±5.60) mm, the distance from the posterior articular surface was (15.01±3.38) mm, the average length of screw was (44.80±3.59) mm. In the CT scan image, all the screws were located in the sustentaculum tali, two cases penetrated the upper bony cortex, no cases showed that screw pierced the inferior and anterior of the sustentaculum tali or lower bony cortex.  Conclusions  The guidance of the self-designed director which adopted the "medial-oriented method" can improve the accuracy of sustentaculum tali screw placement and reduce the complications such as screw entry into the joint and calcaneal medial cortex.

关键词

跟骨 /  载距突 /  导向器 /  解剖 /  影像学

Key words

Calcaneal /  Sustentaculum Tali /  Director /  Anatomical /  Radiographic

引用本文

导出引用
刘金伟, 朱朝晖, 张增方, 刘得恒, 杨斌, 郑良孝, 陈东亮, 宋晓峰, 刘鹏, 赛佳明, 丁自海. 跟骨内侧定位载距突螺钉导向器的研制与初步评价[J]. 中国临床解剖学杂志. 2019, 37(6): 691-696 https://doi.org/10.13418/j.issn.1001-165x.2019.06.016
LIU Jin-wei, ZHU Zhao-hui, ZHANG Zeng-fang, LIU De-heng, YANG Bin, ZHENG Liang-xiao, CHEN Dong-liang, SONG Xiao-feng, LIU Peng, SAI Jia-ming, DING Zi-hai. Development and preliminary evaluation of the calcaneal sustentaculum tali director using the method of positioning from the medial of the calcaneus [J]. Chinese Journal of Clinical Anatomy. 2019, 37(6): 691-696 https://doi.org/10.13418/j.issn.1001-165x.2019.06.016
中图分类号: R683.42   

参考文献

[1]  Wang C, Huang D, Ma X, et al. Sustentacular screw placement with guidance during ORIF of calcaneal fracture: an anatomical specimen study[J]. J Orthop Surg Res, 2017, 12(1): 78.
[2] Qiang M, Chen Y, Zhang K, et al. Effect of sustentaculum screw placement on outcomes of intra-articular calcaneal fracture osteosynthesis: a prospective cohort study using 3D CT[J]. Int J Surg, 2015, 19(1): 72-77.
[3] Bussewitz BW, Hyer CF. Screw placement relative to the calcaneal fracture constant fragment: an anatomic study[J]. J Foot Ankle Surg, 2015, 54(3): 392-394.
[4]  Gitajn IL, Toussaint RJ, Kwon JY. Assessing accuracy of sustentaculum screw placement during calcaneal fixation[J]. Foot Ankle Int, 2013, 34(2): 282-286.
[5] Phisitkul P, Sullivan JP, Goetz JE, et al. Maximizing safety in screw placement for posterior facet fixation in calcaneus fractures: a cadaveric radio-anatomical study[J]. Foot Ankle Int, 2013, 34(9): 1279-1285.
[6]  Olexa TA, Ebraheim NA, Haman SP. The sustentaculum tali: anatomic, radiographic, and surgical considerations[J]. Foot Ankle Int, 2000, 21(5): 400-403.
[7]  梅炯, 俞光荣, 朱辉, 等. 跟骨载距突的解剖特点及其临床意义[J]. 中国临床解剖学杂志, 2002, 20(1): 9-11.
[8]  张翅, 袁仕国, 周孟君, 等. 点对点定位器辅助置入固定跟骨骨折载距突螺钉的初步报告[J]. 中华创伤骨科杂志, 2010, 12(11): 1095-1097.
[9] De Boer AS, Van Lieshout EMM, Vellekoop L, et al. 2D and 3D assessment of sustentaculum tali screw fixation with or without screw targeting clamp[J]. Injury, 2017, 48(12): 2864-2871.
[10] Geerling J, Kendoff D, Citak M. Intraoperative 3D imaging in calcaneal fracture care-clinical implications and decision making[J]. J Trauma, 2009, 66(3): 768-773.
[11] 刘丙根, 庞清江, 余霄, 等. 自行研制的跟骨载距突螺钉导向器辅助固定跟骨关节内骨折[J]. 中华创伤骨科杂志, 2015, 17(9): 786-790.
[12] 张耘, 赵金忠. Broden位X线监控技术在跟骨骨折手术治疗中的应用[J]. 中华骨科杂志, 2004, 24(1): 36-38.
[13] 王志杰, 邹云雯, 钟世镇, 等. 载距突的解剖学研究及其在跟骨关节内移位骨折手术中的应用[J]. 中华创伤骨科杂志, 2009, 11(2): 129-132.
[14] Gras F, Marintschev I, Wilharm A, et al. Sustentaculum tali screw placement for calcaneus fractures-different navigation procedures compared to the conventional technique[J]. Z Orthop Unfall, 2010, 148(3): 309-318.
[15] Burke CJ, Adler RS. Tibial nerve block using an ultrasound-guided inframalleolar medial plantar nerve perineural injection: a technical note[J]. J Clin Ultrasound, 2017, 45(3): 134-137.

基金

青岛市医疗卫生重点学科建设项目(QDZDXK-A-2017005);山东大学齐鲁医院(青岛)科研启动基金资助项目(QDKY2018QN01)