股骨远端解剖数据测量及其与膝关节假体径线的匹配性探讨

薛瑞琪,许杰,翁健豪,朱琪琪,李登,蔡志清,黄玉麟,马若凡

中国临床解剖学杂志 ›› 2015, Vol. 33 ›› Issue (6) : 631-636.

中国临床解剖学杂志 ›› 2015, Vol. 33 ›› Issue (6) : 631-636. DOI: 10.13418/j.issn.1001-165x.2015.06.004
应用解剖

股骨远端解剖数据测量及其与膝关节假体径线的匹配性探讨

  • 薛瑞琪1,2, 许杰1, 翁健豪1,2, 朱琪琪1,2, 李登1, 蔡志清1, 黄玉麟1, 马若凡1
作者信息 +

Comparison between the morphology of Chinese people’s distal femurs and the parameters of knee joint prostheses

  • XUE Rui-qi 1,2, XU Jie 1, WENG Jian-hao 1,2, ZHU Qi-qi 1,2, LI Deng 1, CAI Zhi-qing 1, HUANG Yu-lin1, MA Ruo-fan1
Author information +
文章历史 +

摘要

目的 将股骨远端解剖结构相关数据与国内常用的膝关节假体参数进行对比,评估假体吻合度,为设计符合解剖结构的假体提供参考。  方法 ①选100例无明显骨质缺损的华南地区成人股骨标本 (左51,右49), 拍数字化照片;②校正放大率后测量:机械轴长度、机械轴/解剖轴夹角、髁横径、髁间截骨面横径、外髁前后径、内髁前后径;③测量Zimmer NexGen LPS、Stryker Scrorpio NRG Knee-Flexed、United-U1、威高 GKPS假体的髁内外径宽、假体前髁高度及后髁高度。   结果 (1)股骨远端横径为(7.60±0.60) cm,股骨远端外侧髁高度为(5.90±0.76) cm,内侧髁高度为(5.73±0.74) cm,其间无明显相关性;(2) 4品牌假体的截骨长度均可适应股骨远端骨结构,但前髁截骨高度与4种品牌假体前髁高度对比,在设定容差范围内的涵盖率为24%~55%,股骨后髁高度与假体的后髁高度对比,涵盖率仅14%~21%。   结论 股骨远端解剖形态数据分布较为弥散,在关节置换过程中需个体化处理。假体径线与华南地区国人股骨远端形态存在一定差异。

Abstract

Objective In the present study we compared the morphological parameters of distal femurs with the relevant parameters of knee prostheses, intending to obtain data for better-fit prosthesis design. Methods (1) One hundred adult femur samples from South China (left 51, right 49) with no evident bony defect are selected and photographed. (2) The magnification is corrected according to the photographed scale. Data of measured femoral mechanical axis lengths, angles between mechanical axis/anatomical axis, transverse diameters of the condyles, transverse diameters of distal condylar osteotomic surfaces, anteroposterior diameters of the medial and lateral condyles were obtained. (3) Meanwhile, the widths between the internal and external condylar, heights of the anterior and posterior condyles of the prostheses (both domestic and imported including Zimmer NexGen LPS, Stryker Scrorpio NRG Knee-Flexed, United-U1 and GKPS) were measured. Results (1) The average distal femoral transverse diameter was (7.60±0.60) cm; the average height of the lateral condyles was (5.90±0.76) cm; the average height of the medial condyles was (5.73±0.74) cm. No evident relativity existed e between these data. (2)The osteotomic lengths of the prostheses from all four brands fitted Chinese people’s distal femur. By comparing the ostotomic heights of anterior condyles and the heights of anterior prosthetic condyle, a coverage rate of 14%~21% was obtained which fell within the range of tolerance. Similarly, we got a coverage rate of only 24%~55% posteriorly. Conclusions Since the morphological data of Chinese people’s distal femur are dispersive, individualized replacement is required. The parameters of prostheses do not completely match the morphology of distal fumurs of Chinese people.

关键词

  / 假体 / 股骨远端 / 膝关节 / 测量

Key words

 Prosthesis / Distal femur / Knee / Measurement

引用本文

导出引用
薛瑞琪,许杰,翁健豪,朱琪琪,李登,蔡志清,黄玉麟,马若凡. 股骨远端解剖数据测量及其与膝关节假体径线的匹配性探讨[J]. 中国临床解剖学杂志. 2015, 33(6): 631-636 https://doi.org/10.13418/j.issn.1001-165x.2015.06.004
Comparison between the morphology of Chinese people’s distal femurs and the parameters of knee joint prostheses[J]. Chinese Journal of Clinical Anatomy. 2015, 33(6): 631-636 https://doi.org/10.13418/j.issn.1001-165x.2015.06.004

参考文献

[1] Dalury DF, Mason JB, Murphy JA, et al. Analysis of the outcome in male and female patients using a unisex total knee replacement system[J]. J Bone Joint Surg Br, 2009,91(3): 357-360.
[2]  Xie F, Lo NN, Pullenayegum EM, et al. Evaluation of health outcomes in osteoarthritis patients after total knee replacement: a two-year follow up[J]. Health Quality Life Outcomes, 2010,8: 87. 
[3]  Ritter MA, Wing JT, Berend ME, et al. The clinical effect of gender on outcome of total knee arthroplasty[J]. Arthroplasty,  2008, 23(3): 331-336.
[4]  Erkman MJ, Walker PS. A study of knee geometry applied to the design of condylar prosthesis[J].Biomed Eng, 1974, 9(1): 14- 17.
[5]  Cheng CK, Lung CY, Lee YM, et al. A new approach of designing the tibial baseplate of total knee prosthesis[J]. Clin Biomech, 1999, 14(2): 112-117.
[6]  Hofmann AA, Evanich JD, Ferguson RP, et al. Ten- to 14-year clinical followup of the cementless Natural Knee system[J]. Clin Orthop Relat Res. 2001, 388:85-94.  
[7]  Freeman MA, Todd RC, Bamert P, et al. ICLH arthroplasty of  the knee: 1968--1977[J]. J Bone Joint Surg Br. 1978, 60-B(3): 339-344.  
[8] Insall IN. Revision of aseptic failed total knee arthroplasty[A]//W. Norman Scott. Insall & Scott Surgery of the Knee. New York:Churchill Livingston,1993:935-957.
[9]  张建, 董纪元, 付忠田. 人膝关节尺寸与5种人工膝关节假体尺寸的对照[J].中国组织工程研究与临床康复, 2009, 13(4):635-638.
[10] 李梁,吴海山.国人股骨远端形态测量在全膝关节置换与假体设计中的意义[J].中国组织工程研究, 2013, 17(43):7521-7526.
[11] Kirby Hitt MD. Anthropometric measurements of the human knee:correlation to the sizing of current knee arthroplasty systems[J]. J Bone Joint Surg, 2003, 85A(Suppl 4):115-122.
[12]朱志刚, 金大地, 罗吉伟,等. 股骨远端解剖形态性别差异与假体形态设计的关系[J].中国临床解剖学杂志, 2008, 26(4):386-388.

基金

广东省科技计划项目(2014A020215009)